| Catalog Number |
RBY2C0520 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Physical Resistance (2578)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure treating an arteriovenous malformation (avm) using ruby coils.During the procedure, the physician experienced resistance while manipulating the ruby coil within the lantern delivery microcatheter (lantern) and the coil unintentionally detached in the target location.The procedure was completed with only this ruby coil.It was reported that the physician used an rhv but did not maintain continuous flush.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note the device is no longer available for return as mentioned in the initial mdr; therefore, the mdr was updated accordingly.The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
|
| |
|
Search Alerts/Recalls
|