Model Number N/A
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
International customer reported that the check-flo valve of a peripherally inserted central catheter (picc) is loose.The customer indicated that the issue can result in potential infection and an inability to fully administer treatment due to possible leakage.There are no reported adverse patient consequences associated with this report.The handling conditions and circumstances surrounding the event are not known.No further information was provided.No product will be returned for evaluation.
 
Additional Manufacturer Narrative
A review of the complaint history, device history record, instructions for use (ifu), quality control, and specifications of the device was conducted during the investigation, and no issues were found related to the reported issue.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT STANDARD POWER INJECTABLE PICC LINE PICC
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6427179
Report Number1820334-2017-00533
Device Sequence Number11038400
Product Code LJS
UDI-Device Identifier00827002037885
UDI-Public(01)00827002037885(17)170621(10)5974159
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUPICDS-5.0-CT-NT-1110
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/21/2017
Supplement Date Received by Manufacturer01/24/2018
Initial Report FDA Received Date03/23/2017
Supplement Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-