• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION GLIDESHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION GLIDESHEATH; INTRODUCER, CATHETER Back to Search Results
Catalog Number 50-1060
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 02/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual device was not returned to the manufacturing facility for evaluation.Therefore the investigation for this report is limited.A review of the device history record of the product code/lot# combination was conducted with no relevant findings.Two other devices were used in this same event and were also reported to leak, only one actual sample was returned.See mdr number 1118880-2017-00013 for the sample evaluation.There is no evidence that this event was related to a device defect or malfunction.The reported issue could not be confirmed without the complaint sample.The exact cause of the reported event cannot be definitively determined based on the available information.All available information has been placed on file in quality assurance at the manufacturing facility for appropriate tracking, trending and follow-up.Actual device not returned.
 
Event or Problem Description
The user facility reported valve leakage during a procedure.Follow up communication with the user facility confirmed the following information: a backflow of blood was reported coming out of the valve from the side port after it is flushed out; blood loss was less than 250cc; the procedure was completed and the outcome was reported as good; there was no patient injury or medical intervention required, patient is stable; and two other devices were used in this same event and were also reported to leak, see mdr numbers 1118880-2017-00013 and 1118880-2017-00014.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key6428043
Report Number1118880-2017-00015
Device Sequence Number10772282
Product Code DYB
UDI-Device Identifier00389701011332
UDI-Public(01)00389701011332(17)180831(10)UG22
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K142183
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue Number50-1060
Device Lot NumberUG22
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2017
Initial Report FDA Received Date03/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CATHETERS; WIRES
Patient Age61 YR
-
-