MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH; PERCUTANEOUS, IMPLANTED, LONG TERM INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH; PERCUTANEOUS, IMPLANTED, LONG TERM INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 0600330
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/22/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leaking chronic catheter was confirmed and the cause appears to be use related.The product returned for evaluation was one 9fr d/l hickman catheter.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated and a split was observed just distal of the bifurcation, on the white-lured lumen side.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: 'c' shaped split.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).The catheter material was visibly lighter in color at the fracture site.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.Reference (b)(4) for further information about this failure type.The product nursing guide provides the following guidance pertaining to occluded or partially occluded catheters: ¿to avoid catheter rupture, do not force entire amount into catheter if strong resistance is felt.¿ a lot history review (lhr) is not possible since a valid lot number was not provided by the complainant.
 
Event or Problem Description
As reported by the facility, the hickman catheter was found broken at one of the extension legs.The catheter was then handled and removed according to hospital protocol.No harm to patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH
Common Device Name
PERCUTANEOUS, IMPLANTED, LONG TERM INTRAVASCULAR CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key6428453
Report Number3006260740-2017-00294
Device Sequence Number10842292
Product Code LJS
UDI-Device Identifier00801741051753
UDI-Public(01)00801741051753
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0600330
Device Lot NumberHUYH006 (INVALID)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Was the Report Sent to FDA? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2017
Initial Report FDA Received Date03/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-