Catalog Number AB14W030150150
Device Problems Burst Container or Vessel (1074); Material Fragmentation (1261)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 02/23/2017
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
During an attempt to treat the posterior tibial artery (pta) reported to be severely calcified with little tortuosity using nanocross elite balloon catheter, it was reported that the nanocross elite balloon ruptured on first inflation.After the balloon ruptured the physician could not get the balloon back out of the vessel.The balloon then detached from the balloon catheter.Physician made numerous attempts to snare and remove the balloon.Remaining portion of the balloon catheter was removed and the sheaths were removed.The physician was not able to retrieve the balloon and remains in the patient.No further clinical sequelae reported for this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOCROSS 0.14 OTW PTA DILATATION CATHETER
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key6428972
MDR Text Key70698646
Report Number2183870-2017-00140
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00821684066508
UDI-Public00821684066508
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/26/2019
Device Catalogue NumberAB14W030150150
Device Lot NumberA218242
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2017
Date Device Manufactured01/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-