• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTLAB AIRFLOW; RESUS BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTLAB AIRFLOW; RESUS BAG Back to Search Results
Model Number AF5140MB
Device Problem Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/24/2017
Event Type  malfunction  
Manufacturer Narrative
An extensive and thorough investigation was performed by the engineering team that concluded the root cause for this problem is the fact that the patient port and the mask are made out of similar base materials resulting in an undesirably high coefficient of friction between the two parts.Steps have been taken to implement a material combination with a lower coefficient of friction that significantly reduces the effort required to remove the mask from the patient valve.
 
Event Description
The customer alleges "mask is difficult to remove." no other details were provided and no patient injury/harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRFLOW
Type of Device
RESUS BAG
Manufacturer (Section D)
VENTLAB
2710 northridge dr.
suite a
grand rapids MI 49544
Manufacturer (Section G)
VENTLAB
2710 northridge dr.
suite a
grand rapids MI 49544
Manufacturer Contact
carrie fortuna
2710 northridge dr.
suite a
grand rapids, MI 49544
6162598400
MDR Report Key6429227
MDR Text Key70909955
Report Number2246980-2017-00016
Device Sequence Number1
Product Code BTM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 02/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberAF5140MB
Device Lot Number305700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2017
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage N
Patient Sequence Number1
-
-