This is filed to report that the steerable guide catheter (sgc) soft tip was torn.It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3 and at the same time to treat the tricuspid valve.Echo imaging was challenging.Two clips were successfully implanted, reducing mr to <1.To treat the tricuspid valve, the mitraclip delivery system (cds) (61129u123) was advanced to the tricuspid valve, however during positioning the clip got caught in the chordae, causing tissue damage.The clip was freed from the chordae and the decision was made to discontinue the procedure.The clip was in the closed position when retracting into the sgc, and the clip got caught in the guide tip of the sgc.After removal of the devices it was noted that the gripper line was broken, the grippers could not be raised.A tear was noted on the soft tip of the sgc.No clips were implanted in the tricuspid valve.There were no adverse patient effects and there was no clinically significant delay during the procedure.The patient is stable.No additional treatment is planned.No additional information was provided.
|
(b)(4).The reported device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All available information was investigated and the reported tears in the soft tip resulting from the clip getting caught on the guide tip appears to be related to procedural conditions/user technique.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|