Catalog Number IGTCFS-65-1-JUG-TULIP |
Device Problem
Difficult or Delayed Positioning (1157)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Investigation is still in progress.
|
|
Event Description
|
Description of event according to initial reporter: the filter wouldn't deploy from the hook.After some manipulation the filter came off the hook and placement was as intended.Patient outcome: the patient did not require any additional procedures due to this occurrence.No adverse effects to the patient due to this occurrence.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Summary of investigational findings: investigation is solely based on description of event, no product was returned and no imaging was provided to assist the investigation.Given the limited information provided, it would be inappropriate to speculate at what may or may not have led to the reported filter leg not expanding after deployment.No evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
|
|
Search Alerts/Recalls
|