All in-process specifications and release criteria were met, including pre-and-post-sterile seal strength testing on the packaging, fourier-transform infrared spectroscopy and coating density testing on the cured coated c-qur mosaic panel, and visual inspection of the coating of the cut sample.Ball burst, suture retention (course and wale directions), and ftir testing were conducted on the polypropylene mesh material at incoming and all acceptance criteria was also met.Clinical evaluation: c-qur mosaic mesh is a sterile, knitted polypropylene monofilament mesh material with a non-continuous coating derived from biological oil composed of fatty acids, lipids and glycerides (bao bioabsorbable oil coating).This product is intended for use in soft tissue deficiencies including hernia repair, chest wall reconstruction, traumatic or surgical wounds and other fascial surgical intervention procedures requiring reinforcement with a supportive material.The bao coating is absorbable by the body.If in the event the coating does flake or come off during a procedure it would cause no harm to the patient, the mesh would then become the equivalent of bare mesh.The instructions for use state under adverse reaction, that complications may occur with the use of any surgical mesh include, but are not limited to, mechanical disruption of the tissue and/or mesh material, possible adhesions when placed in direct contact with the viscera (intestines) and organs.
|