• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIGHT SCIENCES, INC. VISCO360 VISCOSURGICAL SYSTEM; VISCOELASTIC INJECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIGHT SCIENCES, INC. VISCO360 VISCOSURGICAL SYSTEM; VISCOELASTIC INJECTOR Back to Search Results
Model Number 03916
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All pertinent information available to sight sciences, inc.Has been submitted.The company is submitting this mdr to ensure full compliance with 21 cfr 803.(b)(4).Submitted to fda on: 03/24/2017.
 
Event or Problem Description
The visco360 device was used on (b)(6) 2017 to perform viscodilation of schlemm's canal on the right eye.After fully retracting the catheter from the first 180 degrees of dilation of schlemm's canal, the surgeon noticed a very small blue fragment entangled in the iris (presumably from the microcatheter).The surgeon was able to complete the procedure and viscodilate the remaining 180 degrees of the canal, but elected to leave the fragment in the iris due to concerns about potential complications involved in its removal.On (b)(6) 2017 surgeon reported the patient's post-operative iop was 8 mmhg with some hyphema.On (b)(6) 2017, the surgeon reported that the patient was healing well with an iop of 11 mmhg (unmedicated).He was able to view the fragment under high magnification and estimated it was 10-20 microns.The fragment was observed in the same area where the trabecular meshwork was accessed to perform viscodilation.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).Submitted to fda on: 03/28/2017.
 
Additional Manufacturer Narrative
Additional data provided in response to letter dated may 10, 2017.This supplemental report is being re-submitted since the prior attempt to submit this information failed.Mfr reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISCO360 VISCOSURGICAL SYSTEM
Common Device Name
VISCOELASTIC INJECTOR
Manufacturer (Section D)
SIGHT SCIENCES, INC.
3000 sand hill road, 3-105
menlo park CA 94025
MDR Report Key6431427
Report Number3010363671-2017-00002
Device Sequence Number16333616
Product Code MRH
UDI-Device Identifier00858027006013
UDI-Public0100858027006013
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143205
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/31/2017
Device Model Number03916
Device Catalogue Number03916
Device Lot Number1005191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/27/2017
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/24/2017
Supplement Report FDA Received Date03/28/2017
06/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-