Model Number HEWROOT0030
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/23/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that on (b)(6) 2017, during a cardiac surgery, the involved graft frayed.The surgeon was creating a bio bentall with the involved graft and a magna heart valve.The surgeon reported that it necessitated stopping twice to deal with the fraying of the graft.According to the surgeon, the patient was quite ill with ejection fraction 25% and the added cardiac ischemic time and cardiopulmonary bypass (cpb) time related to this graft issue was severely suboptimal.In the surgeon's estimation it extended his cardiac ischemic period of his already compromised heart by 10-15 minutes and also his cpb time if the graft had frayed any more.The surgery was completed with the involved graft, and no adverse outcome for the patient was reported.
 
Additional Manufacturer Narrative
(b)(4) the outcome of the investigation would tend to indicate that the product was not defective at the time of manufacturing (no failure detected).However, due to their intrinsic weaving pattern, woven grafts are prone to fraying.Considering the reported event (frayed graft at the time of use : out of specification), a non-conformance report has been initiated in order to investigate the root cause and take appropriate corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOROOT AORTIC GRAFT
Common Device Name
VASCULAR POLYESTER GRAFT
Manufacturer (Section D)
INTERVASCULAR SAS
zi athelia 1
la ciotat cedex, 13705
FR  13705
MDR Report Key6431450
Report Number1640201-2017-00006
Device Sequence Number16278699
Product Code DSY
UDI-Device Identifier00384401013914
UDI-Public00384401013914
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103347
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/30/2021
Device Model NumberHEWROOT0030
Device Catalogue NumberHEWROOT0030
Device Lot Number16G21
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/24/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/24/2017
Supplement Report FDA Received Date04/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-