Model Number A2-80072
Medical Device Problem Codes Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 03/15/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer is still waiting for the arrival of the affected iv sets.
 
Event or Problem Description
The user reported that the iv sets were leaking from bottom of drip chamber.No patient was involved in this case.
 
Additional Manufacturer Narrative
The manufacturer received the investigation report from its contract supplier on 09/12/2017.The leaking issue was confirmed.The root cause was due to inadequate adhesion process performed between the drip chamber and the hose.The re-training of the operation personnel at the contract supplier site has been enforced on 05/16/2017 to address similar reported issues.(b)(4).
 
Event or Problem Description
This is a follow-up for the initially filed mdr (3006575795-2017-00073).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key6432712
Report Number3006575795-2017-00073
Device Sequence Number29613
Product Code FPA
UDI-Device Identifier00814371020112
UDI-Public00814371020112
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2-80072
Device Lot Number1604008
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2017
Supplement Date Received by Manufacturer03/15/2017
Initial Report FDA Received Date03/24/2017
Supplement Report FDA Received Date09/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-