| Model Number |
A2-80072 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
03/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer is still waiting for the arrival of the affected iv sets.
|
| |
|
Event or Problem Description
|
|
The user reported that the iv sets were leaking from bottom of drip chamber.No patient was involved in this case.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer received the investigation report from its contract supplier on 09/12/2017.The leaking issue was confirmed.The root cause was due to inadequate adhesion process performed between the drip chamber and the hose.The re-training of the operation personnel at the contract supplier site has been enforced on 05/16/2017 to address similar reported issues.(b)(4).
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2017-00073).
|
| |
|