| Model Number |
B2-70071-P |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
03/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The failed iv sets have been sent to the contract supplier for evaluation on 03/17/2017.The manufacturer is still waiting for the results.
|
| |
|
Event or Problem Description
|
|
The user reported that leakage of an iv set occurred past the roller clamp."this time i had tested pumps and the tube left hanging, the bag only had about 375ml remaining-which all leaked out past the 2 closed clamps".No patient was involved in this case.
|
| |
|
Additional Manufacturer Narrative
|
|
Zyno medical has received the investigation report from the contract manufacturer on 04/25/2017.The user-reported complaint was confirmed.The root cause was unknown and was found to not be related to the manufacturing process.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795- 2017-00075).
|
| |
|