ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number B2-70071-P
Medical Device Problem Codes Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 03/10/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The failed iv sets have been sent to the contract supplier for evaluation on 03/17/2017.The manufacturer is still waiting for the results.
 
Event or Problem Description
The user reported that leakage of an iv set occurred past the roller clamp."this time i had tested pumps and the tube left hanging, the bag only had about 375ml remaining-which all leaked out past the 2 closed clamps".No patient was involved in this case.
 
Additional Manufacturer Narrative
Zyno medical has received the investigation report from the contract manufacturer on 04/25/2017.The user-reported complaint was confirmed.The root cause was unknown and was found to not be related to the manufacturing process.
 
Event or Problem Description
This is a follow-up for the initially filed mdr (3006575795- 2017-00075).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key6432746
Report Number3006575795-2017-00075
Device Sequence Number18963111
Product Code FPA
UDI-Device Identifier00814371020181
UDI-Public00814371020181
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB2-70071-P
Device Lot Number16105861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/10/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/24/2017
Supplement Report FDA Received Date05/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-