• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Nonstandard Device (1420); Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/28/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Recall: 1627487-05242011-002-r, 1627487-07262012-002-r.This ipg serial number was included in field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the patient's scs ipg was not functioning.The ipg was determined to be inoperable.Subsequently, the ipg was replaced.Stimulation therapy was reportedly restored postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6434487
MDR Text Key70857111
Report Number1627487-2017-01508
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 02/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2011
Device Model Number3788
Device Lot Number2891001
Was Device Available for Evaluation? No
Date Manufacturer Received02/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-12192011-003-R
Patient Sequence Number1
Treatment
MODEL 1194 (X2), SCS ANCHOR; MODEL 3219, SCS LEAD
Patient Outcome(s) Other;
Patient Age50 YR
-
-