TERUMO CARDIOVASCULAR SYSTEMS CORP. LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 801041 |
Device Problem
Material Distortion (2977)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/03/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The srt replaced the drive belt.Roller pump meets manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the drive belt from the roller pump was stretched affecting the drive torque.There was no patient involvement.
|
|