| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause of malfunction: user's test strip had poor storage (bathroom).(b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 297mg/dl.The customer's expected fasting blood glucose test result range is 120 to 130mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the bathroom.The test strip lot manufacturer's expiration date is 02/28/2019 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history (date / time not set): (b)(6).
|
| |
|