• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ENCORE HUMERAL SHOULDER STEM; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ12X175MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ENCORE HUMERAL SHOULDER STEM; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ12X175MM Back to Search Results
Catalog Number 530-12-175
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 02/28/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having a non-union fracture that never healed.Loose implant.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had a fracture that never healed and the implant loosened.The in-vivo length of patient service for the implant was 2.1 years.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had a fracture that would not heal and the implant loosened.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE HUMERAL SHOULDER STEM
Common Device Name
ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ12X175MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6437312
Report Number1644408-2017-00177
Device Sequence Number541181
Product Code PHX
UDI-Device Identifier00888912168434
UDI-Public(01)00888912168434
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/22/2022
Device Catalogue Number530-12-175
Device Lot Number413T1005
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/28/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/27/2017
Supplement Report FDA Received Date06/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-32-103,LOT 864C1917; 509-00-432,LOT 385P1056
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-