ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES BI-PHASIC
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to the lithium battery on the system board.The device operator's guide instructs users to replace the lithium battery on the device every 5 years.This device is approximately 17 years old.Analysis for reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device was intermittently unable to power on.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6437396
Report Number1220908-2017-00601
Device Sequence Number18911384
Product Code MKJ
UDI-Device Identifier00847946003609
UDI-Public00847946003609
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K990762
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2017
Initial Report FDA Received Date03/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-