| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "defib fault 80" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed during testing.However, the reported problem could not be duplicated.The device was put through extensive testing without duplicating the malfunction.The device's defib flex cable and pacer flex cable were replaced as a precaution.No trend is associated with reports of this type.
|
| |
|