| Catalog Number |
65SP035 |
| Medical Device Problem Code |
Air Leak (1008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event date: (b)(6) 2017.Pma/510(k) number: k944178 and k083641.Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.
|
| |
|
Event or Problem Description
|
|
It was reported that a portex® bivona® aire-cuf® neonatal and pediatric tracheostomy tube air in the cuff decreased to half-capacity 6 hours after initial use with a patient.The issue was observed by medical personnel.It was unknown if the tube had been checked prior to use.The tracheostomy tube was replaced with a new one.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One bivona® 3.5mm aire-cuf® neonatal and pediatric tracheostomy tube was returned for investigation.During functional testing, a syringe was used to inflate the device cuff and inflation balloon with 5cc's of air and the device was submerged under water; the device was manipulated while under the water.No bubbles (indicating a leak) were found escaping from the device.The cuff was removed from the water and the device cuff was inflated with 5cc's of air and allowed to sit for 4 hours.A loss of volume of the cuff was observed; however, a leak could not be detected.Investigation was unable to determine the root cause of the reported issue; however, no evidence was found to suggest a manufacturing defect.
|
| |
|