• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ANTHEM CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ANTHEM CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3112
Device Problem Incorrect Measurement (1383)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/23/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Upon interrogation, the device showed false eri resulting in explant and replacement.The patient is in stable condition.
 
Manufacturer Narrative
The complaint of early battery depletion could not be confirmed.The device was received in normal working condition, with battery voltage above eri level.Analysis indicated a false eri trip has occurred, and the date shown as (b)(4) 2013 which is before the manufacturing date.False eri trip could happen when the device exposed to strong emi.In those cases, clearing the eri message with the programmer restores normal function.Electrical and mechanical tests indicated that the device exhibited normal characteristics with no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6440656
MDR Text Key71092463
Report Number2017865-2017-01841
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2015
Device Model NumberPM3112
Device Lot Number4306290
Other Device ID Number05414734502832
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2017
Initial Date FDA Received03/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-