• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA SPINE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA SPINE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA SPINE 1.0.2
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2016
Event Type  malfunction  
Manufacturer Narrative
The device sp 14 003 has been inspected for investigation purpose.The tests performed confirmed that the screw thread was damaged.As repair, the damaged part were replaced.The internal complaint reference is the following: (b)(4).
 
Event Description
It has been reported that during a surgery, the patient reference screw was bloused during the patient registration.A surgery delay < 30 minutes has been observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SPINE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
7414400
MDR Report Key6441961
MDR Text Key71101098
Report Number3009185973-2017-00185
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K151511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA SPINE 1.0.2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2016
Initial Date FDA Received03/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-