| Model Number |
MT22430-BLU |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.No product or data were provided for evaluation.The reported no audio output could not be confirmed.A root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.The device was visually inspected and no defects were found.A manual test was performed and speaker sounded.The reported event of no audio output was not confirmed.A root cause could not be determined.
|
| |
|
Search Alerts/Recalls
|