| Model Number |
ZM-541PA |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reports that the transmitter got hot and melted the battery compartment.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Further evaluation being performed.
|
| |
|
Event or Problem Description
|
|
The customer reports that the transmitter got hot and melted the battery compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer narrative: the customer reports that the transmitter got hot and melted the battery compartment.The unit was cleaned and evaluated.The reported problem of getting hot was not duplicated.However, it was found that the spo2 connector was broken and the non-invasive blood pressure (nibp) function failed.All malfunctioning parts have been replaced.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 24 hours of extended testing and operates to manufacturer's specifications.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|
| |
|