Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, air ingress was observed during aspiration.The balloon catheter was reinserted into the sheath without resolution.The balloon catheter was replaced as the physician was concerned that the shaft might have been deformed.The replacement balloon catheter was inserted into the sheath and air ingress was continued to be observed.The sheath was then replaced with temporary resolution.When aspiration was performed in the latter half of the procedure, however, a small amount of air was observed.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files for the date of the event were returned and analyzed.The data files showed eight injections performed with balloon catheter, 2af284 lot 83206, without issue.The reported air ingress cannot be confirmed through data analysis; results are pending investigation of the returned physical product.
 
Manufacturer Narrative
Product event summary: the sheath was returned and analyzed.Visual inspection of the sheath showed the device was intact with no apparent issues.Air aspiration was reproduced when a test balloon catheter was introduced through the sheath.Dissection showed the hemostatic valve was leaking; the valve was torn.The reported issue (air ingress) has been confirmed through testing.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6447184
MDR Text Key71325277
Report Number3002648230-2017-00140
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 03/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/11/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number59823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-