• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS SENSAR; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS SENSAR; MONOFOCAL IOLS Back to Search Results
Model Number AR40E
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a scratch was found on an intraocular lens (iol) before surgery, prior to folding the lens.No patient involvement or injury was reported.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 04/12/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed scratches on the lens optic.The customer's reported complaint was verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6447457
Report Number2648035-2017-00603
Device Sequence Number406104
Product Code HQL
UDI-Device Identifier05050474502017
UDI-Public(01)05050474502017(17)210521
Combination Product (Y/N)N
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/21/2021
Device Model NumberAR40E
Device Catalogue NumberAR40E00155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/02/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/30/2017
Supplement Report FDA Received Date05/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-