• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, a 980 ventilator failed the power on self test (post) and generated an inoperable error message.The ventilator was not in use on a patient at the time the event occurred.The service engineer (se) inspected the device and found a diagnostic code relevant to the reported incident recorded in the ventilators memory logs.The se replaced the power supply.The se updated the software to the current revision, performed calibrations and extended self-testing on the device and all tests passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key6448594
Report Number8020893-2017-05499
Device Sequence Number13561242
Product Code CBK
UDI-Device Identifier10884521171633
UDI-Public10884521171633
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K131252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2017
Initial Report FDA Received Date03/30/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-