| Catalog Number |
9734887 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Skin Irritation (2076); Tissue Damage (2104); Iatrogenic Source (2498)
|
| Date of Event |
03/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information: per further engineering review, the reported event was hazard coded with a minor harm code.Review of the biocompatibility testing of the device found that the device passed the irritation index.No further investigation shall occur at this time.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that the non-invasive patient tracker produced "skin damage" on the patient's forehead.The patient was in a prone position for the seven hour surgery.The rep postulated that the patient tracker was not making direct contact with the horseshoe headholder, but was unable to visually confirm this during the surgery.The system performed as expected throughout the case.The skin damage was identified one day after the surgery and the hospital was unable to specify what type of skin damage occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: per further engineering review, the reported event was hazard coded with a minor harm code.Review of the biocompatibility testing of the device found that the device passed the irritation index.No further investigation shall occur at this time.
|
| |
|
Search Alerts/Recalls
|