• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. SIMULUS SEMI-RIGID ANNULOPASTY RING; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. SIMULUS SEMI-RIGID ANNULOPASTY RING; RING, ANNULOPLASTY Back to Search Results
Model Number 800SR40
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/13/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the device has not been received for analysis.Without the receipt of the product, no definitive conclusion can be made regarding the clinical observation.Conclusion: surgeons often attempt to repair mitral valves in lieu of replacing them due to improved long term clinical outcomes.There are times when a valve repair is attempted using an annuloplasty device and subsequent post repair evaluation demonstrates an inadequate results.This is often due to suboptimal anatomy, surgical technique, or inappropriate sizing, and not a malfunction of the device.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received additional information from this physician that this device was implanted and explanted in the same procedure because the ring was too small.The device was replaced with another annuloplasty ring.There were no ring related problems, and no other adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMULUS SEMI-RIGID ANNULOPASTY RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
MDR Report Key6452031
Report Number3008592544-2017-00014
Device Sequence Number16362345
Product Code KRH
UDI-Device Identifier00643169182233
UDI-Public00643169182233
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
D125551
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number800SR40
Device Catalogue Number800
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/23/2017
Supplement Date Received by Manufacturer03/23/2017
Initial Report FDA Received Date04/02/2017
Supplement Report FDA Received Date10/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient Weight94
-
-