| Model Number |
MSERIES CCT |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing the device displayed a "defib fault 72" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed and attributed to a system interconnection cable that was not properly seated.The cable was replaced to correct the reported problem.Analysis of reports of this type has not identified an increase in trend.
|
| |
|