Model Number N/A
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).We received one used, half filled easypump ii lt 125-25-s without packaging.The provided sample was subjected to a visual inspection.Damages were not detected.As-received condition the clamp clip was closed and the patient connector was not closed (the original wing cap was not handed over by the customer).Further on, we detected no liquid and crystallized drug residues at the filling port (lli-cone) and at the patient connector (lla-cone) of the sample.Additionally the pump was filled with nacl 0.9 % up to the nominal value (125 ml) and a functional test respectively a leak test was carried out.After starting the pump the pump did work immediately (solution was running).Leakages were not detected.Furthermore the flow rate of the pump was tested.Nominal: 5 ml/h actual: 7.8 ml in 1 h; 14.9 ml in 2 hrs; 83.8 ml in 16 hrs.The flow rate of the inspected pump is within our specifications.A too fast flow (as described by the customer) could not be determined.Device history records (dhr): reviewed device history records, there is no abnormalities and no such defect detected during in process and at final control inspection.
 
Event or Problem Description
As reported by the user facility (translation of user facility information by bbm sales organization in france): pump-flow rate-fast.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP® II
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key6454000
Report Number9610825-2017-00064
Device Sequence Number13797493
Product Code MEB
UDI-Device Identifier04046964448454
UDI-Public(01)04046964448454(17)210501(10)16F14GE211
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K081905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/06/2017
Report Date (Section F) 04/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2021
Device Model NumberN/A
Device Catalogue Number4540006
Device Lot Number16F14GE211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2017
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer04/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2017
Initial Report FDA Received Date04/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA04/03/2017
-
-