MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 X-FORCE BALLOON DILATION CATHETER, 7MM X 10 CM, STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 X-FORCE BALLOON DILATION CATHETER, 7MM X 10 CM, STERILE Back to Search Results
Catalog Number 997710
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the balloon leaked.It was later reported that the procedure was performed to treat a ureteral stricture.The catheter balloon was dilated at the ureter of the patient.There was no resistance when inserting the catheter.The balloon catheter was placed and dilated at the ureter for about two minutes, after which the balloon catheter began to leak.The procedure; however, was completed without any further reported problems.
 
Additional Manufacturer Narrative
Received 1 used x-force dilation catheter.The reported issue was confirmed; however, the cause is unknown.Upon return of the sample, a visual inspection did not note any obvious defects.The balloon was intact with no visible tears or missing pieces.The shaft of the catheter did not show any kinks and appeared intact.The catheter was submerged in water and inflated with air using an in-house eagle inflation device.Testing noted that air leaked from the inflation arm near the hub.Further examination noted the presence of glue at the connection of the hub to the inflation arm which indicated that it was a full glue joint.The glue was cured and was hard.There was no evidence of any glue discrepancies.The root cause for this complaint could not be determined; however, it could be due to mishandling of the device or excessive force applied to the hub during the procedure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: " description the x-force® balloon dilation catheter is a dual lumen catheter with a balloon mounted on the distal tip.It has a radiopaque tip and two radiopaque markers beneath the balloon that define the working length.The lumen labeled with the rated burst pressure (xx atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038" (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure, typically 10 atm.The x-force® balloon dilation catheter comes packaged with a refolding tool and is available with or without an inflation device.It comes sterile and is for single patient use only.Indications for use: the x-force® balloon dilation catheter is recommended for use in the dilation of the urinary tract.Contraindications: do not use the x-force® balloon dilation catheter in the presence of conditions, which create unacceptable risk during the dilation of the urinary tract.Warnings: ¿ do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, noncompliant tissue or in the presence of certain calculi.¿ this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: caution: federal law (u.S.A.) restricts this device to sale by or on the order of a physician.Only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device.After use, the x-force® balloon dilation catheter and/or accessories may be considered a potential biohazard.Handle and dispose of in accordance with medical practice and applicable laws and regulations.Inspection prior to use: the x-force® balloon dilation catheter is a sterile, single patient use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Preparation of the catheter: all x-force® balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.Remove the protective sheath from the balloon.Attach the inflation device to the connector on the balloon lumen.Open the stopcock and draw back on the inflation device to remove the air from the balloon catheter.Close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.Repeat steps 3-4 until all air is removed from the balloon lumen.Catheter insertion: introduce the catheter carefully over a 0.038" (.97mm) guidewire and place it in the area that needs to be dilated.Use the radiopaque markers to aid in proper positioning.Note: dilation procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment or direct vision.Caution: do not advance or withdraw the catheter or guidewire against any significant resistance.The cause of the resistance must be determined fluoroscopically and remedial action taken.Inflating the balloon catheter: fill the inflation device (eagle¿ inflation device) with sterile diluted contrast medium.Attach the inflation device to the balloon lumen.Inflate the balloon note: do not use air or any gaseous substances as a balloon inflation medium always use sterile liquid media.Note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the rated burst pressure of the balloon is not exceeded.Caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to re-inflate the balloon.If after inspection it is noted pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible.Warning: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, noncompliant tissue or in the presence of certain calculi.Deflating the balloon catheter: deflate the balloon using an inflation device.Since deflation times vary based on balloon sizes and shaft lengths, check fluoroscopically to confirm deflation before attempting to withdraw.Attach the inflation device (eagle¿ inflation device) to the balloon catheter and remove the solution from the balloon by pulling back on the inflation device.Remove the inflation device from the balloon catheter.This will allow ambient pressure to enter, relaxing the balloon.Gently withdraw the catheter.Use of a gentle counterclockwise twisting motion is recommended when removing the catheter.Caution: if resistance is felt when removing either the catheter or the guidewire from the introducer sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.Using the refolding tool: the balloon catheter is supplied with a refolding tool.This aids in preparing the balloon for another insertion during the procedure.To use: manually compress the balloon.Position the refolding tool at one end of the balloon.Twist the refolding tool counter clockwise and push down on the balloon until the refolding tool traverses the entire length of the balloon.Once the balloon is folded, remove the refolding tool and proceed with the procedure.Dispose of properly.Storage: store the balloon dilation catheter in a cool, dry, dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." (b)(4).
 
Event or Problem Description
It was reported that the balloon leaked.It was later reported that the procedure was performed to treat a ureteral stricture.The catheter balloon was dilated at the ureter of the patient.There was no resistance when inserting the catheter.The balloon catheter was placed and dilated at the ureter for about two minutes, after which the balloon catheter began to leak.The procedure; however, was completed without any further reported problems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-FORCE BALLOON DILATION CATHETER, 7MM X 10 CM, STERILE
Common Device Name
X-FORCE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6455009
Report Number1018233-2017-01520
Device Sequence Number591797
Product Code EZN
UDI-Device Identifier00801741052149
UDI-Public(01)00801741052149(17)170630(10)BMYGE007
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Catalogue Number997710
Device Lot NumberBMYGE007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2017
Initial Date Received by Manufacturer 03/13/2017
Supplement Date Received by Manufacturer06/24/2017
Initial Report FDA Received Date04/03/2017
Supplement Report FDA Received Date07/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-