| Catalog Number |
48811118 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
03/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; 18mm plate was implanted & when trying to tighten screws down surgeon was unable to get pass spring bar.End was bent not allowing screw to engage.Plate was removed and new selected with rescue screw used.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection, device history review, complaint history review, risk assessment.The device was not confirmed to have a deformed spring bar and functional testing revealed a blocker able to be turned.Manufacturing files were reviewed and no anomalies were found.Potential root causes include:-surgeon does not tighten screws properly; - screw hole not properly prepared (too shallow).
|
| |
|
Event or Problem Description
|
|
It was reported that; 18mm plate was implanted & when trying to tighten screws down surgeon was unable to get pass spring bar.End was bent not allowing screw to engage.Plate was removed and new selected with rescue screw used.
|
| |
|