• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52916
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Name and address for importer site: (b)(4).Investigation is still in progress.
 
Event Description
Description of event according to initial reporter: "filter would not detach from the delivery system in the body.They recovered the filter and removed it." patient outcome: the patient did not require any additional procedures due to this occurrence.No adverse effects to the patient due to this occurrence.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Summary of investigational findings: the jugular introducer and long dialator are returned.The sheath tip is slightly damaged - probably because the filter was pulled back.The grasping hook is compared to the a test device and no anomalies are noted.The root cause for the reported failed detachment cannot be determined.No evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov DK-46-32
DA   DK-4632
56868686
MDR Report Key6455736
MDR Text Key71587419
Report Number3002808486-2017-00870
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529165
UDI-Public(01)10827002529165(17)170319(10)E3195629
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG52916
Device Catalogue NumberIGTCFS-65-1-JUG-TULIP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Distributor Facility Aware Date03/17/2017
Device Age36 MO
Initial Date Manufacturer Received 03/15/2017
Initial Date FDA Received04/04/2017
Supplement Dates Manufacturer Received03/17/2017
Supplement Dates FDA Received08/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-