Model Number G52916 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Name and address for importer site: (b)(4).Investigation is still in progress.
|
|
Event Description
|
Description of event according to initial reporter: "filter would not detach from the delivery system in the body.They recovered the filter and removed it." patient outcome: the patient did not require any additional procedures due to this occurrence.No adverse effects to the patient due to this occurrence.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Summary of investigational findings: the jugular introducer and long dialator are returned.The sheath tip is slightly damaged - probably because the filter was pulled back.The grasping hook is compared to the a test device and no anomalies are noted.The root cause for the reported failed detachment cannot be determined.No evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
|
|