Model Number ROSA 2.5.8
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
The device (b)(4) has been inspected for investigation purpose.The tests performed confirmed that hydraulic stabilisation system was damaged and an oil leak below the actuator was also identified.The left front hydraulic actuator was fixed by unlocking the nut screw previously locked and a readjustment of the levelling foot was performed as repair purpose.The internal complaint reference is the following: (b)(4).
 
Event Description
It has been reported that during a surgery, the hydraulic stabilisation system was non-functional.There was no patient/user impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
7414400
MDR Report Key6456771
MDR Text Key71629119
Report Number3009185973-2017-00469
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-