Model Number ROSA 2.5.8
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
The device (b)(4) has been inspected for investigation purpose.The tests performed confirmed that the fixation process steps of the mayfield adaptor was not properly followed and damaged the fixation of the adapter on the support arm.The damaged components were replaced for repair purpose.The internal complaint reference is the following: (b)(4).
 
Event Description
It has been reported that during a surgery, the mayfield adapter was hard to screw and the fixation was not correct.No patient impact has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
7414400
MDR Report Key6456809
MDR Text Key71628446
Report Number3009185973-2017-00470
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2016
Initial Date FDA Received04/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-