| Model Number |
ACT1520 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the incident sample has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the temperature was not displayed after puncturing.There was no patient harm.Another device was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
One act1520 was received for evaluation.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.Visual inspection found no defects.The customer reported that the temperature was not stable.The reported condition was confirmed.The water circulated normally through the sample but the sample did not read the temperature properly.Evaluation found the thermocouple solder joint was broken.This is a manufacturing related failure.Corrective action was implemented subsequent to the manufacture of the incident device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported the temperature was not displayed after starting the ablation.There was no patient harm.Another device was used to complete the procedure.
|
| |
|