Catalog Number 5C5479
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Event date: event occurred on an unspecified date in 2017.The device was received and is in the process of being evaluated.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an amia automated pd cycler set was warped, would not lay flat, and would not fit in an amia cycler.This was observed prior to patient connection.The patient proceeded with a new cassette.There was no damage to the cassette packaging.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
One cassette and heater bag were received for evaluation.Visual inspection identified a warped cassette, and a damaged (torn) heater bag.Functional testing performed included leak testing, and clear passage test with no issues noted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was determined to be a deformed cassette.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMIA AUTOMATED PD CYCLER SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6457070
MDR Text Key71687979
Report Number1416980-2017-02974
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/23/2019
Device Catalogue Number5C5479
Device Lot NumberH16I23047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/10/2017
Initial Date FDA Received04/04/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/17/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
AMIA CYCLER
-
-