BAXTER HEALTHCARE - MOUNTAIN HOME AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number 5C5479 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).Event date: event occurred on an unspecified date in 2017.The device was received and is in the process of being evaluated.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an amia automated pd cycler set was warped, would not lay flat, and would not fit in an amia cycler.This was observed prior to patient connection.The patient proceeded with a new cassette.There was no damage to the cassette packaging.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
One cassette and heater bag were received for evaluation.Visual inspection identified a warped cassette, and a damaged (torn) heater bag.Functional testing performed included leak testing, and clear passage test with no issues noted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was determined to be a deformed cassette.Should additional relevant information become available, a supplemental report will be submitted.
|
|