The customer contacted a siemens customer care center (ccc) specialist.The ccc was granted remote access.The ccc reviewed the data provided.The customer's quality control (qc) was in at the time of the event.The ccc did not find any process errors associated with the event.The customer performed patient comparisons on their alternate dimension vista instruments, resulting within the expected clinical range.The ccc performed qc, resulting within range.The ccc performed a full quickcheck, failing for probe leak.The customer reseated the tubing and the ccc performed the reagent prep probe and saw no obvious leaks.The ccc primed the prop probe 30 times and repeated the quickcheck, resulting within specifications.A siemens customer service engineer (cse) was dispatched to the customer's site.The cse replaced the sample probe 2 mixer.The cse aligned the sample probe 2 probe.The cse performed an overmix test, resulting within specifications.The cse performed and quickcheck and qc, resulting within range.The cause of the discordant, falsely elevated total bilirubin results is due to a malfunction of the sample probe 2 mixer.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely elevated total bilirubin results were obtained on patient samples on a dimension vista 1500 instrument.The initial results were reported out to the physician(s).The customer repeated the same samples on an alternate dimension vista instrument, resulting lower.The customer issued corrected reports to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated total bilirubin results.
|