Suspect medical device brand name = pipeline flex w/shield technology model # = ped2-350-25 the device was returned for evaluation and the clinical observation could not be confirmed.As received the pipeline flex with shield delivery system was stuck within the distal segment of the microcatheter.During evaluation, the pipeline with shield delivery system was pushed out of from the catheter with difficulty and found intact.No separation was observed with the pushwire.The distal and proximal ends of the pipeline braid were found fully opened with slightly frayed.Additionally, the pushwire found bent at the proximal end.All other subassembly found normal.All products are 100% inspected for damage and irregularities during manufacture.It is possible that the reported moderate vessel tortuosity may have contributed to the high friction during resheathing; subsequently causing the pipeline flex with shield to become damaged.Additionally, it was reported that the physician continued to resheath the pipeline flex with shield despite of high friction.Per our instructions for use (ifu), the user should: ¿discontinue delivery of the device if high force or excessive friction is encountered during delivery.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intraluminal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.
|