• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L XTRA AUTOTRANSFUSION SYSTEM, PROCEDURE SET BX/225; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L XTRA AUTOTRANSFUSION SYSTEM, PROCEDURE SET BX/225; APPARATUS, AUTOTRANSFUSION Back to Search Results
Catalog Number 04264
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was not provided.The age of the device was calculated as the time elapsed between the device sterilization and the date of the event.(b)(4).Sorin group (b)(4) manufactures the xtra autotransfusion system, procedure set bx/225.The incident occurred in (b)(6).Per exemption number e2016005.The failed connection was a connection made by the user during set-up of the device on the autotransfusion equipment.Sorin group (b)(4) has determined that this issue was caused by user error.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Device not available for return.
 
Event or Problem Description
Sorin group (b)(6) received a report that the blue tubing from the xtra procedure set disconnected from the reservoir during a case and blood spilled on floor and doctor.The tubing was reconnected and the procedure was completed without further issue.There was no patient injury, and no injury was reported for the doctor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTRA AUTOTRANSFUSION SYSTEM, PROCEDURE SET BX/225
Common Device Name
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L
strada statale 12 nord, 86
mirandola (modena) 41037
IT  41037
MDR Report Key6459594
Report Number9680841-2017-00011
Device Sequence Number563828
Product Code CAC
UDI-Device Identifier08033178110808
UDI-Public(01)08033178110808(17)191114(10)1611140164
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K101586
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2019
Device Catalogue Number04264
Device Lot Number1611140164
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/07/2017
Device Age4 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2017
Initial Report FDA Received Date04/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-