• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-505
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/03/2017
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the involved cartridge confirmed the venous connector was broken.The lot was released for distribution having met all product design, quality and product released criteria.No defects were found during incoming inspection of this lot.Nxstage medical considers this report closed.No additional information will be provided.
 
Event Description
A report was received on march 6, 2017 regarding a critical care patient where the venous tip of the luer broke off in the patient¿s catheter.There was no reported adverse event and no report that medical intervention was required.
 
Event Description
Additional information was received on april 10, 2017.A (b)(6) male patient was receiving crrt on (b)(6) 2016 and when the nurse attempted to disconnect the venous line from the patient's access the venous port broke inside the patient's catheter.The patient's dialysis catheter was removed as the patient remained stable and was discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc
350 merrimack street
lawrence, MA 01843
9786874700
MDR Report Key6459798
MDR Text Key72178338
Report Number3003464075-2017-00021
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133547
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2018
Device Model NumberCAR-505
Device Catalogue NumberCAR-505
Device Lot Number60578001
Other Device ID Number+M535CAR5050/$$021860578001X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/06/2017
Initial Date FDA Received04/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age57 YR
Patient Weight95
-
-