MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22430
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.The receiver was returned for evaluation.An exterior visual inspection was performed and passed.A manual test was performed and failed.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key6462386
Report Number3004753838-2017-29558
Device Sequence Number16284707
Product Code MDS
UDI-Device Identifier10386270000078
UDI-Public(01)10386270000078(241)STK-KD-001(10)5159486
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22430
Device Catalogue NumberSTK-KD-001
Device Lot Number5159486
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2017
Initial Report FDA Received Date04/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3004753838-02/29/16-001C
Patient Sequence Number1
Patient Age14 YR
-
-