NUVASIVE, INC. NUVASIVE PRECEPT SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE, Back to Search Results
Model Number 8800000
Device Problems Detachment Of Device Component (1104); Unintended Movement (3026)
Patient Problem Failure of Implant (1924)
Event Date 03/13/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information or radiographs have been received confirming the event type.Initial surgery date is unknown, as is the interval from implantation to the event.Material identification other than material number is not known.Review of manufacturing records indicate no non-conformances have been identified related to the material number.No revision is known to have been planned/completed.No patient injury has been reported.No root cause has been determined at this time.Labeling review: potential adverse events and complications: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation ".Warnings, cautions and precautions: ".Confirm heights of screws match with patient's lordosis when securing lock screws.Failure to verify screw height following insertion may result in inadequate rod normalization." ".Final-tighten all lock screws with the counter-torque and torque t-handle.Do not final-tighten through other instruments in the set, as the rod may not be able to normalize to the tulip.Ensure there is enough rod overhang when final tightening, and do not lock down on the conical portion of the rod.Failure to do so may lead to improper lock down of the construct.It may be necessary for the surgeon to add length to the rod, depending upon patient anatomy and desired lordosis." not returned and no information supplied.
 
Event Description
On 03/13/2017 nuvasive was notified of a post-operative event in which posterior fixation component(s) were reported to have loosened.No patient injury has been reported.No interventions to correct the construct fault are known to have occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PRECEPT SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE,
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
george panfili
7475 lusk blvd
san diego, CA 92121
8588825019
MDR Report Key6464899
MDR Text Key72402400
Report Number2031966-2017-00038
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517344076
UDI-Public00887517344076
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number8800000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-