ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR Back to Search Results
Model Number E SERIES
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported malfunction was observed and attributed to the lithium battery on the system board.Zoll medical corporation recommends replacement of the lithium battery on the system board every 5 years.This device is approximately 9 years old from date of manufacture.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device was unable to power on.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
E SERIES
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6464937
Report Number1220908-2017-00732
Device Sequence Number579855
Product Code MKJ
UDI-Device Identifier00847946009717
UDI-Public00847946009717
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2017
Initial Report FDA Received Date04/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-