• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number A2-80071-DF120
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer is still waiting for the arrival of the iv sets.
 
Event or Problem Description
The user reported a tubing seperation issue of the iv set during infusion.The tubing disconnected from the filter house.No patient harm or injury was involved in this case.
 
Additional Manufacturer Narrative
The user reported issue was not duplicated using the unused samples from the same lot number.The failed iv set was not captured by the user, but samples from the same lot number were sent back to zyno medical for evaluation.The test result was available on 04/13/2017.
 
Event or Problem Description
This is a follow-up for the initially filed mdr (3006575795-2017-00090).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key6472422
Report Number3006575795-2017-00090
Device Sequence Number364541
Product Code FPA
UDI-Device Identifier00814371020082
UDI-Public00814371020082
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2-80071-DF120
Device Lot Number16037122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/22/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/07/2017
Supplement Report FDA Received Date04/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-