Model Number ZCT150
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implant and explant dates: if implanted or explanted, give date: not applicable no patient contact reported.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that during handling but prior to insertion, foreign material was noticed on an intraocular lens (iol) under a microscope.Reportedly, the material could not be removed.No patient contact was reported.No further information was provided.
 
Additional Manufacturer Narrative
(b)(4).Device available for evaluation? yes.Returned to manufacturer on: 04/21/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 12x microscope magnification showed that the lens was cut in half, most probably to make the removal possible.The customer's reported complaint for foreign material on the lens was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens¿ optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6472800
Report Number9614546-2017-00282
Device Sequence Number18973068
Product Code HQL
UDI-Device Identifier05050474552203
UDI-Public(01)05050474552203(17)200622
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/22/2020
Device Model NumberZCT150
Device Catalogue NumberZCT150U230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/13/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/07/2017
Supplement Report FDA Received Date06/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-