|
Approximately 91 cm of the peritoneal catheter was returned.The returned catheter was patent.However, it did not meet the requirements for leak testing due to a tear observed at the 2 cm from the proximal end.The catheter met the requirements for the tensile strength and elongation tests.The instructions for use that accompany the device cautions that low tear strength is a characteristic of most unreinforced silicone elastomer materials, and that care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks, or tears.A review of the manufacturing records showed no anomalies.All catheters are 100% inspected at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|