| Model Number |
420189-09 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure mode.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to the reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the failure mode were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a cable at the distal tip of the double fenestrated grasper instrument was observed to be frayed.There was no report that any fragment(s) from the instrument fell into a patient during the surgical procedure and there was no report of patient harm, adverse outcome or injury.The planned surgical procedure was completed.
|
| |
|