U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PANORAMIC CORPORATION PANORAMIC; PC 4000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PANORAMIC CORPORATION PANORAMIC; PC 4000 Back to Search Results
Medical Device Problem Codes Maintenance Does Not Comply To Manufacturers Recommendations (2974); Power Problem (3010)
Health Effect - Clinical Codes Electric Shock (2554); No Known Impact Or Consequence To Patient (2692)
Date of Event 03/14/2017
Type of Reportable Event Serious Injury
Event or Problem Description
The (b)(6) tech called in and said one of the staff members, (b)(6), got zapped.After calling the office and speaking to her, she stated that she went to power down the machine and felt the buzz.She said it was like when a child sticks their finger in a wall outlet.She had slight tingling in her arm for a couple hours.She also mentioned that they made a report at the office as well.She said she's fine.She just told her boss what happened and they called their tech in to investigate.Found out the power switch at the base of the machine was broken and per a previous case on (b)(6) 2016 this has been broken for some time.Quoted parts cost but the office never did the repair.Now someone got shocked while trying to press the switch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANORAMIC
Common Device Name
PC 4000
Manufacturer (Section D)
PANORAMIC CORPORATION
4321 goshen road
fort wayne IN 46818
MDR Report Key6475535
Report Number1832462-2017-00006
Device Sequence Number849825
Product Code EHD
UDI-Device Identifier00862309000347
UDI-Public00862309000347
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K043557
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dental Assistant
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2017
Initial Report FDA Received Date04/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-